Low-noise rotating sample holder for ultrafast transient spectroscopy at cryogenic temperatures

Back to all publications

Publication date
DOI http://dx.doi.org/10.1063/1.2735557
Reference R. Fanciulli, J. Cerjak and J. Herek, Low-noise rotating sample holder for ultrafast transient spectroscopy at cryogenic temperatures, Rev. Sci. Instrum. 78, (Article number: 53102), 1-5 (2007)

We present the design and testing of a rotating device that fits within a commercial helium cryostat and is capable of providing at 4 K a fresh sample surface for subsequent shots of a 1–10 kHz amplified pulsed laser. We benchmark this rotator in a transient-absorption experiment on molecular switches. After showing that the device introduces only a small amount of additional noise, we demonstrate how the effect of signal degradation due to high fluence is completely resolved.