A fast position sensitive microstrip-gas-chamber detector at high count rate operation

Back to all publications

Publication date
Reference I.P. Dolbnya, H.P. Alberda, F.G. Hartjes, F. Udo, R.E. Bakker, M. Konijnenburg, E. Homan, J. Cerjak, P. Goedtkindt and W. Bras, A fast position sensitive microstrip-gas-chamber detector at high count rate operation, Rev. Sci. Instrum. 73, 3754-3758 (2002)

Testing of a newly developed position sensitive high count rate microstrip gas chamber (MSGC) detector at high count rate operation has been carried out at the Dutch-Belgian x-ray scattering beamline at the European Synchrotron Radiation Facility (Grenoble, France) with a high intensity x-ray beam. The measurements show local count rate capabilities up to similar to -4.5×105 counts/s/channel. Experimental data taken with this detector are also shown. These tests show that both time resolution down to 1.5 ms/frame and a reliable operation at high counting rates can be achieved.